PRODUCT SPECIFICATIONS
Enclosure
Material |
NEMA 3R Fiberglass wall mount with two 3/4” conduit entries |
Dimensions: |
11” (H) X 10” (W) X 6.375” (D) (27.9cm X 25.4cm X 16.2cm) |
Weight: |
6.6 lbs (3.0 kg) |
Temperature: |
-4° to 104°F (-20° to 40°C) |
Humidity: |
5-95% RH, non-condensing for indoor or outdoor locations. |
Indicators and Controls
External: |
Flowmeter, green power LED and red fault/low flow LED |
Internal: |
On-Off switch, voltage out adjust and flow rate adjust |
Outputs: |
Two 24 VDC power terminals, SPDT fault/low flow relay contact |
Power In/Out: |
85–264 VAC, 47-63 Hz, 1.2 Amps; 24 VDC, 1.1 Amp max. |
Approvals
Hazloc |
FM approved for Class 1, Division 2 Groups C & D Rated for Sampling from a Class 1 Division 1 Area Groups C & D |
DC Supply: |
UL60950-1, UL508, UL1310(3), EN60950-1, CE Mark |
Pump: |
Diaphragm type rated at 1.0 lpm @ 40” H2O at pressurized leak rate of < 1.0 inch wc drop in 5 seconds at 25 inches wc |
Wetted parts: |
Polycarbonate, Neoprene, Tygon 2075, Silicone Silastic, 304/316ss, Buna-N, Brass, PVC, Glass, Acrylic and User Tubing |
Caution
Do not use for Acid gases such as Cl2, ClO2, HCl, HF, HCN, NH3, NO2, SO2 and others since the PPM gases are absorbed by the tubing, moisture and various wetted parts slowing or completely eliminating response to the target gas. Do not attempt to use below freezing without heat tracing and a means to keep the system above 32⁰ F (0⁰ C). Excessive condensation will occur when sampling from a higher to a lower temperature.
Ordering Information
Pumped Sample Draw with 24 VDC power supply |
821-0231-01 |
Aspirated Sample Draw with 24 VDC power supply |
821-0232-01 |
Close Nipple 1/8 NPT fitting |
550-1010-01 |
Coalescing filter and Close Nipple 1/8 NPT |
821-0233-01 |